老带新
Corticosterone Assay (1 Kit)
1/4

Corticosterone Assay (1 Kit)

分享
品牌: R&D Systems
pdf 下载产品说明书
pdf 下载SDS
用小程序,查商品更便捷
收藏
对比对比
咨询咨询
产品介绍
产品介绍
产品信息
产品详情

Product Summary

The Parameter Corticosterone assay is a 3.5 hour competitive enzyme immunoassay designed to measure Corticosterone in cell culture supernates, serum, plasma, and urine.

Precision

Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of kit components

Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Urine

Intra-Assay PrecisionInter-Assay Precision
Sample123123
n202020202020
Mean (ng/mL)1.243.988.051.394.088.21
Standard Deviation0.0780.1790.6050.0990.2410.462
CV%6.34.57.57.15.95.6

Recovery

The recovery of Corticosterone spiked to levels throughout the range of the assay in various matrices was evaluated.

Sample TypeAverage % RecoveryRange %
Cell Culture Media (n=4)114107-124
Human EDTA Plasma (n=4)10887-127
Human Heparin Plasma (n=4)10797-122
Human Serum (n=4)11092-130
Mouse EDTA Plasma (n=4)10496-111
Mouse Serum (n=4)9679-113
Rat EDTA Plasma (n=4)9877-114
Rat Serum (n=4)9987-108
Urine (n=4)10592-121

Linearity

To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Corticosterone were diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.
Multi-species Corticosterone ELISA Linearity
Multi-species Corticosterone ELISA Linearity
Multi-species Corticosterone ELISA Linearity

Scientific Data

Multi-species Corticosterone ELISA Standard Curve

Assay Procedure

Refer to the product for complete assay procedure.

Bring all reagents and samples to room temperature before use. It is recommended that all samples, standards, and controls be assayed in duplicate.
  1.   Prepare all reagents, standard dilutions, and samples as directed in the product insert.
  2.   Remove excess microplate strips from the plate frame, return them to the foil pouch containing the desiccant pack, and reseal.

  3. 50 µL Primary Antibody Solution
  4.   Add 50 µL of Primary Antibody Solution to all wells except the non-specific binding (NSB) wells. Cover with a plate sealer, and incubate at room temperature for 1 hour on a horizontal orbital microplate shaker.
  5.   Aspirate each well of the microplate and wash, repeating the process 3 times for a total of 4 washes.

  6. 100 µL Pretreatment F
  7.   Add 100 µL of Pretreatment F to all wells.

  8. 50 µL Standard, Control, or Sample
  9.   Add 50 µL of Standard, control, or sample to the appropriate wells.

  10. 50 µL of Calibrator Diluent
  11.   Add 50 µL of the Calibrator Diluent to the NSB and zero standard (B0) wells.

  12. 50 µL Conjugate
  13.   Add 50 µL of Conjugate to all wells. Cover with a new plate sealer, and incubate at room temperature for 2 hours on the shaker.
  14.   Aspirate and wash 4 times.

  15. 200 µL Substrate Solution
  16. Add 200 µL Substrate Solution to each well. Incubate at room temperature for 30 minutes on the benchtop. PROTECT FROM LIGHT.

  17. 100 µL Stop Solution
  18. Add 100 µL of Stop Solution to each well. Read at 450 nm within 30 minutes. Set wavelength correction to 540 nm or 570 nm.

Corticosterone Parameter Assay Kit Summary

Format
96-well strip plate
Assay Length
3.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (50 uL), EDTA Plasma (50 uL), Heparin Plasma (50 uL), Urine (10 uL)
Sensitivity
0.047 ng/mL
Assay Range
0.1 - 25 ng/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Urine)
Specificity
Corticosterone in human, mouse, and rat samples.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
背景
别名
17-Deoxycortisol,C21H30O4,Corticosterone
背景

Background: Corticosterone

Corticosterone, also known as Kendall's compound B and Reichstein substance B, is a glucocorticoid secreted by the adrenal gland cortex. It is produced in response to adrenal cortex stimulation by adrenocorticotropic hormone (ACTH) and is the precursor of aldosterone. Stress increases corticosteroid production. Corticosterone level is measured as an indicator of stress in patients with dietary restrictions, long-term memory impairment, and burn injuries.

Alternate Names:
17-Deoxycortisol;C21H30O4; Corticosterone
研究领域
制备和贮存
保存方式

Preparation and Storage

Shipping
The product is shipped at ambient temperature. Upon receipt, store it immediately at the temperature recommended below.
Storage
Store the unopened product at 2 - 8 °C. Do not use past expiration date.
声明 :本官网所有报价均为常温或者蓝冰运输价格,如有产品需要干冰运输,需另外加收干冰运输费。
货号:
KGE009
一键复制
询价
1Kit
选择数量
当前规格1件起购 
配送至
预计2-3周送达,快递: 免运费,若需干冰额外收费